IMMULITE 2500
Report
- Report Number
- 2247117-2010-00027
- Event Type
- Other
- Date Received
- August 27, 2010
- Date of Event
- August 2, 2010
- Report Date
- August 10, 2010
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC.
- Product Code
- JJQ
- PMA / PMN Number
- K905215
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A SIEMENS HEALTHCARE TECHNICAL SERVICE ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA. ANALYSIS OF THE INSTRUMENT DATA INDICATES THAT THE CAUSE OF THE DISCORDANT TROPONIN RESULTS IS UNK. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT HIGH IMMULITE 2500 TROPONIN RESULTS WERE OBTAINED WITH TWO (2) OF THE THREE (3) SAMPLES DRAWN FROM ONE (1) PT. THE RESULT OF THE FIRST SAMPLE WAS POSITIVE. PER THE LABORATORY'S PROTOCOL, A POSITIVE RESULT MUST BE CONFIRMED WITH A SECOND SAMPLE, WHICH WAS DRAWN AND TESTED NEGATIVE. A THIRD SAMPLE WAS DRAWN AND TESTED POSITIVE. ALL THREE SAMPLES WERE LATER RE-TESTED AND THE RESULTS WERE ALL NEGATIVE (I.E., <0.2 NG/ML). PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NOT REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RESULTS.
A SIEMENS HEALTHCARE TECHNICAL SERVICE ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA. ANALYSIS OF THE INSTRUMENT DATA INDICATES THAT THE CAUSE OF THE DISCORDANT TROPONIN RESULTS IS UNK. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT HIGH IMMULITE 2500 TROPONIN RESULTS WERE OBTAINED WITH TWO (2) OF THE THREE (3) SAMPLES DRAWN FROM ONE (1) PT. THE RESULT OF THE FIRST SAMPLE WAS POSITIVE. PER THE LABORATORY'S PROTOCOL, A POSITIVE RESULT MUST BE CONFIRMED WITH A SECOND SAMPLE, WHICH WAS DRAWN AND TESTED NEGATIVE. A THIRD SAMPLE WAS DRAWN AND TESTED POSITIVE. ALL THREE SAMPLES WERE LATER RE-TESTED AND THE RESULTS WERE ALL NEGATIVE (I.E., <0.2 NG/ML). PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NOT REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RESULTS.
A SIEMENS HEALTHCARE TECHNICAL SERVICE ENGINEER (TSE) ANALYZED THE INSTRUMENT DATA. ANALYSIS OF THE INSTRUMENT DATA INDICATES THAT THE CAUSE OF THE DISCORDANT TROPONIN RESULTS IS UNK. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVAL OF THE DEVICE IS REQUIRED.
DISCORDANT HIGH IMMULITE 2500 TROPONIN RESULTS WERE OBTAINED WITH TWO (2) OF THE THREE (3) SAMPLES DRAWN FROM ONE (1) PT. THE RESULT OF THE FIRST SAMPLE WAS POSITIVE. PER THE LABORATORY'S PROTOCOL, A POSITIVE RESULT MUST BE CONFIRMED WITH A SECOND SAMPLE, WHICH WAS DRAWN AND TESTED NEGATIVE. A THIRD SAMPLE WAS DRAWN AND TESTED POSITIVE. ALL THREE SAMPLES WERE LATER RE-TESTED AND THE RESULTS WERE ALL NEGATIVE (I.E., <0.2 NG/ML). PT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NOT REPORT OF ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMULITE 2500 | IMMUNO-ASSAY | JJQ | SIEMENS HEALTHCARE DIAGNOSTICS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 2 | ||||
| 3 |