FDA Adverse Event Injury Summary report: N

IMPLANTIUM

MDR report key: 1820673 · Received August 26, 2010

Report

Report Number
3005503242-2010-00073
Event Type
Injury
Date Received
August 26, 2010
Date of Event
June 2, 2010
Report Date
July 30, 2010
Manufacturer
DENTIUM USA
Product Code
DZE
PMA / PMN Number
K041368
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: BASED ON INSPECTION AND TEST RESULTS, THE RETURNED PRODUCT HAS BEEN FOUND TO BE WITHIN SPECIFICATION. THEREFORE, THE IMPLANT FAILURE IS MOST LIKELY DUE TO CAUSES OTHER THAN THE PRODUCT SUCH AS SURGICAL MISTAKE, PT BONE CONDITION, PT ORAL HYGIENE, OR PT BEHAVIOR.

Description of Event or Problem · 1

PRODUCT WAS RETURNED AS AN IMPLANT FAILURE BECAUSE OF THE POOR BONE CONDITION. BASED ON THE REPORT, THE PT HAD MODERATE ORAL HYGIENE AND POOR BONE CONDITION. THE FIXTURE WAS PLACED WITH A TRADITIONAL 2 STATE SURGERY IN TOOTH LOCATION #8. NO BONE MATERIAL WAS USED. THE DOCTOR INDICATED THAT THE DEVICE WAS NOT RECEIVED DAMAGED OR DEFECTIVE SUCH THAT IT WOULD NOT PERFORM AS INTENDED. THE PT'S OUTCOME WAS REPORTED AS RECOVERED, NO COMPLICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMPLANTIUM ENDOSSEOUS DENTAL IMPLANT DZE DENTIUM USA FX3410MLC 07K1602

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention