HARMONIC ACE 36 CM
Report
- Report Number
- 1527736-2007-06073
- Event Type
- Malfunction
- Date Received
- September 13, 2007
- Date of Event
- July 23, 2007
- Report Date
- July 25, 2007
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4).EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE A WAS RETURNED WITH THE TISSUE PAD MISSING. AS THE TISSUE PAD WAS NOT RETURNED, FURTHER EVALUATION COULD NOT BE PERFORMED. OUR INSTRUCTION INSERT STATES: CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. THIS CAN RESULT IN POSSIBLE DAMAGE TO THE INSTRUMENT. BOTH CONDITIONS MAY CAUSE A SYSTEM FAILURE SIGNALED BY A CONTINUOUS BEEP WHEN EITHER OF THE FOOT PEDALS IS DEPRESSED AND KEEP THE CLAMP ARM OPEN WHEN BACKCUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. EACH DEVICE IS VISUALLY INSPECTED AND FUNCTIONALLY TESTED PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION AND TESTING. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOUND THAT THE DEVICE B WAS RETURNED WITH THE BLADE SCRATCHED AND CRACKED. DUE TO THE DAMAGE TO THE BLADE, IT WAS CONFIRMED THAT THE DEVICE WAS NON-FUNCTIONAL. THE DEVICE WAS TESTED WITH A GENERATOR AND AN ERROR CODE 5 WAS NOTED. THE IDENTIFIED BLADE DAMAGE MAY HAVE OCCURRED FROM EXTERNAL CONTACT DURING PRE-OP OR GENERAL USE. IN ADDITION, MINOR BLADE DAMAGE MAY INCREASE IN SEVERITY DURING SUBSEQUENT ACTIVATIONS, AND MAY RESULT IN BLADE "LOCKOUT" LATER IN THE PROCEDURE. FOR THIS REASON THE FOLLOWING STATEMENTS WERE INCLUDED IN THE INSTRUCTIONS FOR USE: "AVOID ACCIDENTAL CONTACT WITH ALL METAL OR PLASTIC INSTRUMENTS OR OBJECTS WHEN THE INSTRUMENT IS ACTIVATED. CONTACT WITH STAPLES, CLIPS OR OTHER INSTRUMENTS WHILE THE INSTRUMENT IS ACTIVATED MAY RESULT IN CRACKED OR BROKEN BLADES, WHICH MAY BE IDENTIFIED BY GENERATOR SOLID TONE OR INSTRUMENT ERROR." A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS. RESULTS: (BLADE).
IT WAS REPORTED THAT DURING A LAVH THE DEVICE STOPPED WORKING AND AN ERROR CODE 5 WAS DISPLAYED. NO PT CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HARMONIC ACE 36 CM | LFL | ETHICON ENDO SURGERY, INC (CINCINNATI) | NA | D4HA58 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HANDPIECE| GENERATOR |