FDA Adverse Event Injury Summary report: N

GM HELIX ACQUA IMPLANT,TI,4.3X13

MDR report key: 18205279 · Received November 24, 2023

Report

Report Number
0001222315-2023-029303
Event Type
Injury
Date Received
November 24, 2023
Date of Event
August 18, 2022
Report Date
November 24, 2023
Manufacturer
JJGC S.A.
Product Code
DZE
UDI-DI
07899878024965
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
CA, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN ADA 23. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON 2022-08-18, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: PERI-IMPLANTITIS, PAIN, SWELLING AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186785 GM HELIX ACQUA IMPLANT,TI,4.3X13 ENDOSSEOUS DENTAL IMPLANT DZE JJGC S.A. LGR63 07899878024965

Patients

Seq Age Sex Outcome Treatment
1 50 YR Female Required Intervention