FDA Adverse Event Malfunction Summary report: N

OT PING METER

MDR report key: 1820330 · Received August 30, 2010

Report

Report Number
2939301-2010-07276
Event Type
Malfunction
Date Received
August 30, 2010
Report Date
August 13, 2010
Manufacturer
LIFESCAN, INC.
Product Code
MDS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE LAY USER/PATIENT CONTACTED LIFESCAN ALLEGING THE METER DOES NOT POWER ON WITH TEST STRIP. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WERE NO ALLEGATIONS OF HARM OR INJURY. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OT PING METER GLUCOSE MONITORING SYS/KIT MDS LIFESCAN, INC. 2960357

Patients

Seq Age Sex Outcome Treatment
1