FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL

MDR report key: 1820170 · Received October 30, 2007

Report

Report Number
1527736-2007-07179
Event Type
Malfunction
Date Received
October 30, 2007
Date of Event
September 20, 2007
Report Date
September 26, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DAMAGED JAWS; DROPPING/EJECTING CLIPS. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT (A) WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED 8 CLIPS DUE TO THE CONDITION OF THE JAWS. HOWEVER, NO JAMMING ISSUES WERE NOTED DURING ANALYSIS. THE INSTRUMENT LOCKED OUT AS INTENDED. THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT (B) WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED 13 CLIPS DUE TO THE CONDITION OF THE JAWS. HOWEVER, NO JAMMING ISSUES WERE NOTED DURING ANALYSIS. THE INSTRUMENT LOCKED OUT AS INTENDED. AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE ROOT CAUSE OF THE EJECTED CLIPS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAP CHOLECYSTECTOMY PROCEDURE THERE WERE 2 DEVICES USED, THEY BOTH JAMMED DURING THE FIRING, A THIRD DEVICE WAS USED TO COMPLETE THE CASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL FZP ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA C4GO34

Patients

Seq Age Sex Outcome Treatment
1