LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL
Report
- Report Number
- 1527736-2007-07179
- Event Type
- Malfunction
- Date Received
- October 30, 2007
- Date of Event
- September 20, 2007
- Report Date
- September 26, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED JAWS; DROPPING/EJECTING CLIPS. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT (A) WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED 8 CLIPS DUE TO THE CONDITION OF THE JAWS. HOWEVER, NO JAMMING ISSUES WERE NOTED DURING ANALYSIS. THE INSTRUMENT LOCKED OUT AS INTENDED. THE ANALYSIS RESULTS FOUND THAT THE ER420 INSTRUMENT (B) WAS RETURNED WITH THE JAWS OVER TWISTED. THE INSTRUMENT WAS CYCLED AND EJECTED 13 CLIPS DUE TO THE CONDITION OF THE JAWS. HOWEVER, NO JAMMING ISSUES WERE NOTED DURING ANALYSIS. THE INSTRUMENT LOCKED OUT AS INTENDED. AN INVESTIGATION HAS BEEN INITIATED TO ADDRESS THE ROOT CAUSE OF THE EJECTED CLIPS. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAP CHOLECYSTECTOMY PROCEDURE THERE WERE 2 DEVICES USED, THEY BOTH JAMMED DURING THE FIRING, A THIRD DEVICE WAS USED TO COMPLETE THE CASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIGACLIP ENDOSCOPIC ROTATING MULTIPLE CLIP APPLIER 20 LARGE TITANIUM LIGATING CL | FZP | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | C4GO34 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |