FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1820161 · Received October 30, 2007

Report

Report Number
1527736-2007-07190
Event Type
Malfunction
Date Received
October 30, 2007
Date of Event
August 27, 2007
Report Date
August 27, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC.(CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED WITH A LOOSE MOUNT. IT WAS TESTED ON A GENERATOR AND NO ERROR CODE WAS NOTED. HOWEVER, A HISSING SOUND WAS HEARD. THE HAND PIECE WAS DISASSEMBLED; A TORN ACOUSTIC ISOLATOR WAS FOUND IN THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING ROUTINE HANDPIECE TESTING, THE HANDPIECE HAD A LOOSE MOUNT. NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, INC.(CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR