FDA Adverse Event
Malfunction
Summary report: N
ULTRACISION HARMONIC SCALPEL HAND PIECE
MDR report key: 1820120
·
Received October 30, 2007
Report
- Report Number
- 1527736-2007-07191
- Event Type
- Malfunction
- Date Received
- October 30, 2007
- Report Date
- April 4, 2007
- Manufacturer
- ETHICON ENDO-SURGERY, INC. (CINCINNATI)
- Product Code
- LFL
- PMA / PMN Number
- K002906
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED IN GOOD CONDITION. THE ANALYSIS WAS PERFORMED AND WAS FOUND THAT THE HAND PIECE NO LONGER MEETS THE SPECIFICATIONS FOR IMPEDANCE, PHASE MARGIN. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
Description of Event or Problem · 1
THE DEVICE WAS RETURNED BUT WITHOUT ANY INFORMATION. NO FURTHER INFORMATION REGARDING THIS DEVICE IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRACISION HARMONIC SCALPEL HAND PIECE | LFL | ETHICON ENDO-SURGERY, INC. (CINCINNATI) | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |