FDA Adverse Event Malfunction Summary report: N

ULTRACISION HARMONIC SCALPEL HAND PIECE

MDR report key: 1820120 · Received October 30, 2007

Report

Report Number
1527736-2007-07191
Event Type
Malfunction
Date Received
October 30, 2007
Report Date
April 4, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
LFL
PMA / PMN Number
K002906
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE HAND PIECE WAS RETURNED IN GOOD CONDITION. THE ANALYSIS WAS PERFORMED AND WAS FOUND THAT THE HAND PIECE NO LONGER MEETS THE SPECIFICATIONS FOR IMPEDANCE, PHASE MARGIN. THE HAND PIECE WAS DISASSEMBLED, A TORN ACOUSTIC ISOLATOR AND EVIDENCE OF MOISTURE INGRESS WERE FOUND IN THE INSTRUMENT. COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. THE BATCH HISTORY RECORDS WERE REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED BUT WITHOUT ANY INFORMATION. NO FURTHER INFORMATION REGARDING THIS DEVICE IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ULTRACISION HARMONIC SCALPEL HAND PIECE LFL ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA NA

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR