FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1820115 · Received September 26, 2007

Report

Report Number
1527736-2007-06348
Event Type
Malfunction
Date Received
September 26, 2007
Date of Event
July 10, 2007
Report Date
July 10, 2007
Manufacturer
ETHICON ENDO-SURGERY, INC. (CINCINNATI)
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). JAW OR CAM INTERFACE IS DISENGAGED. EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH JAWS DISENGAGED FROM THE CAM. THE INSTRUMENT WAS CYCLED AND EJECTED THE REMAINING CLIPS DUE TO THE JAWS CONDITION. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE FOUND DAMAGE. WE HAVE DOCUMENTED THE CIRCUMSTANCES AS THEY WERE REPORTED TO US. IN ADDITION, COMPLAINT INFORMATION IS TRENDED ON A REGULAR BASIS TO DETERMINE IF FURTHER INVESTIGATION IS WARRANTED. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE HANDLE DID NOT CLOSE SO THE CLIPS DID NOT COME OUT. ANOTHER DEVICE WAS OPENED AND THE PROCEDURE WAS COMPLETED WITHOUT CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER FZP ETHICON ENDO-SURGERY, INC. (CINCINNATI) NA D4HC3X

Patients

Seq Age Sex Outcome Treatment
1