FDA Adverse Event Malfunction Summary report: N

HUDSON CONCHA NEPTUNE

MDR report key: 1820019 · Received August 25, 2010

Report

Report Number
3004365956-2010-00202
Event Type
Malfunction
Date Received
August 25, 2010
Date of Event
August 13, 2010
Report Date
August 16, 2010
Manufacturer
TELEFLEX MEDICAL
Product Code
BTT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT HAS BEEN RECEIVED BY MANUFACTURER FOR EVALUATION, HOWEVER, INVESTIGATION REPORT IS INCOMPLETE AT THIS TIME. A FOLLOW-UP REPORT WILL BE SENT WHEN ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

THE EVENT IS REPORTED AS: WHILE THE DEVICE WAS IN USE WITH A #780-18 (NEONATAL HEATED DUAL LIMB CIRCUIT), A 3 INCH FLAME APPEARED FROM THE INSPIRATORY LIMB DIRECTLY WHERE THE WIRES EXIT THE CIRCUIT. THE PATIENT HAD TO BE BAGGED WHILE THE CIRCUIT AND HEATER WERE REPLACED. THE INFANT WAS NOT HARMED AND REMAINED ON THE VENTILATOR, AND WAS LISTED IN FINE AND STABLE CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HUDSON CONCHA NEPTUNE HUMIDIFIER BTT TELEFLEX MEDICAL NA NA

Patients

Seq Age Sex Outcome Treatment
1