DARIO BLOOD GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3010606081-2023-00076
- Event Type
- Malfunction
- Date Received
- November 23, 2023
- Date of Event
- October 24, 2023
- Report Date
- October 24, 2023
- Manufacturer
- LABSTYLE INNOVATIONS LTD.
- Product Code
- NBW
- PMA / PMN Number
- K150817
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
DURING TROUBLESHOOTING ON THE PHONE WITH DARIO'S REPRESENTATIVE, IT WAS DETERMINED THAT THE USER WAS STORING HER CARTRIDGE OF STRIPS IN THE BATHROOM. DARIO'S USER GUIDE STATES IN THE SECTION OF GENERAL PRECAUTIONS: "DO NOT STORE THE METER OR TEST STRIPS IN AN AREA WHERE THE HUMIDITY IS OUTSIDE OF THE RANGE 10-90%, SUCH AS A BATHROOM OR KITCHEN." IT WAS ALSO DETERMINED THAT THE USER WAS USING ALCOHOL SWABS TO CLEAN HER FINGER BEFORE TESTING. DARIO'S USER GUIDE STATES IN THE SECTION FACTORS THAT INTERFERE WITH BLOOD GLUCOSE TESTING "ALCOHOL CAN AFFECT TEST RESULTS". A REPLACEMENT OF DARIO'S STRIPS AND CONTROL SOLUTION WAS SENT TO THE USER TO MAKE SURE THAT THE DARIO STRIPS ARE READING CORRECTLY, AND TO ENSURE PROPER TECHNIQUE. AFTER THE NEW DARIO STRIPS AND CONTROL SOLUTION WERE RECEIVED, DARIO'S REPRESENTATIVE FOLLOWED UP WITH THE USER. THE USER'S DARIO METER WAS TESTED WITH CONTROL SOLUTION AND A NEW CARTRIDGE OF STRIPS: THE METER READ WITHIN RANGE ONLY FOR THE LEVEL 2 TEST. CONTROL SOLUTION LEVEL 1 TEST RESULTS WAS 157 MG/DL (RANGE 107-155 MG/DL) CONTROL SOLUTION LEVEL 2 TEST RESULTS WAS 252 MG/DL (RANGE 183-264 MG/DL). THE USER ALSO TESTED HER BG LEVEL AND RECEIVED A BG READING OF 109 MG/DL, WHICH THE USER STATED WAS IN RANGE FOR HER. MULTIPLE ATTEMPTS TO FOLLOW UP WITH THE USER AND RETEST WITH THE CONTROL SOLUTION WERE MADE, HOWEVER, NO RESPONSE HAS BEEN RECEIVED TO DATE. THE HIGH BG READINGS MAY HAVE BEEN DUE TO MISUSE OF THE STRIPS. THERE IS NOT ENOUGH INFORMATION REGARDING THE METER AND OLD STRIPS AVAILABLE TO FURTHER INVESTIGATE THIS CASE. THEREFORE, NO RESOLUTION IS AVAILABLE.
ON (B)(6) 2023, THE USER CONTACTED DARIO TO REPORT HIGH BLOOD GLUCOSE (BG) READINGS.THE USER REPORTED THAT PRIOR TO CONTACTING DARIO, SHE RECEIVED A BG READING OF 170 MG/DL FROM HER DARIO METER COMPARED TO A BG READING OF 103 FROM HER NON-DARIO METER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 230765 | DARIO BLOOD GLUCOSE MONITORING SYSTEM | GLUCOMETER | NBW | LABSTYLE INNOVATIONS LTD. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Other |