FDA Adverse Event Malfunction Summary report: N

BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC)

MDR report key: 18195943 · Received November 22, 2023

Report

Report Number
2647876-2023-00584
Event Type
Malfunction
Date Received
November 22, 2023
Date of Event
October 22, 2023
Report Date
December 8, 2023
Manufacturer
BECTON DICKINSON CARIBE LTD.
Product Code
MDB
UDI-DI
00382904420239
PMA / PMN Number
K113558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G5: THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR THE ADDITIONAL 510K IS AS FOLLOWS: K113558 AND K222591. H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. A SAMPLE IS AVAILABLE FOR EVALUATION. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: CATALOG: 442023. BATCH NO.: 3130646. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) TEN PATIENT ACCESSION NUMBER (B)(6) MOLECULAR FALSE POSITIVE RESULTS WERE RECEIVED BY THE USER WHILE USING IN CONJUNCTION WITH THE BIOFIRE PLATFORM. A DUAL IDENTIFICATION OF C. TROPICALIS AND STREPTOCOCCUS VIRIDANS GROUP WAS RECEIVED. THE C. TROPICALIS WAS THE UNEXPECTED RESULT. THE STREPTOCOCCUS VIRIDANS GROUP RESULT WAS CONFIRMED BY GRAM STAIN AND BACTERIAL CULTURE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 6 OF 10.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) TEN PATIENT (ACCESSION NUMBER (B)(6)) MOLECULAR FALSE POSITIVE RESULTS WERE RECEIVED BY THE USER WHILE USING IN CONJUNCTION WITH THE BIOFIRE PLATFORM. A DUAL IDENTIFICATION OF C. TROPICALIS AND STREPTOCOCCUS VIRIDANS GROUP WAS RECEIVED. THE C. TROPICALIS WAS THE UNEXPECTED RESULT. THE STREPTOCOCCUS VIRIDANS GROUP RESULT WAS CONFIRMED BY GRAM STAIN AND BACTERIAL CULTURE. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. REPORT 6 OF 10.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2042168 BD BACTEC¿ PLUS AEROBIC/F CULTURE VIALS (PLASTIC) SYSTEM, BLOOD CULTURING MDB BECTON DICKINSON CARIBE LTD. 3130646 00382904420239

Patients

Seq Age Sex Outcome Treatment
1 Unknown