BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2023-00580
- Event Type
- Malfunction
- Date Received
- November 22, 2023
- Date of Event
- October 6, 2023
- Report Date
- December 18, 2023
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420215
- PMA / PMN Number
- K123903
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.6 INVESTIGATION SUMMARY: CATALOG: 442021. BATCH NO.: 3152014. CUSTOMER REPORTED A MOLECULAR FALSE POSITIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE THE CUSTOMER¿S EXPERIENCE WITH THE BACTEC PRODUCT BASED ON OUR INTERNAL PROCEDURES AND THE INTENDED USE OF THE PRODUCT. RETENTION SAMPLES WERE VISUALLY INSPECTED, TESTED FOR VIABLE CONTAMINATION BY SUB-CULTURING ON TSA, CHOCOLATE, SABOURAUD AND SCHAEDLER AGARS PLATES, VOLTAGE OUTPUT AND GRAM STAIN. ALL RESULTS WERE SATISFACTORY. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT. A COMPLAINT HISTORY REVIEW WAS CONDUCTED, AND BATCH HAS BEEN PREVIOUSLY INVESTIGATED FOR THE REPORTED DEFECT. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH HISTORY RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. H3 OTHER TEXT : SEE H.10.
H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
REPORT 4 OF 8. IT HAS BEEN REPORTED THAT THE BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) HAS BEEN FOUND GIVING MOLECULAR FALSE POSITIVE RESULTS DURING USE. NO PATIENT IMPACT WAS REPORTED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: REPORTED MOLECULAR FP WHICH WAS C.TROPICALIS WITH BCID 2.
REPORT 4 OF 8 IT HAS BEEN REPORTED THAT THE BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) HAS BEEN FOUND GIVING MOLECULAR FALSE POSITIVE RESULTS DURING USE. NO PATIENT IMPACT WAS REPORTED. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: REPORTED MOLECULAR FP WHICH WAS C.TROPICALIS WITH BCID 2.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1621222 | BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 3152014 | 00382904420215 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |