FDA Adverse Event Injury Summary report: N

HAKIM PERITON CATH,120CM

MDR report key: 18192182 · Received November 22, 2023

Report

Report Number
3013886523-2023-00392
Event Type
Injury
Date Received
November 22, 2023
Date of Event
October 18, 2023
Report Date
January 19, 2024
Manufacturer
INTEGRA LIFESCIENCES MANSFIELD
Product Code
JXG
UDI-DI
10381780518839
PMA / PMN Number
K944222
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE HAKIM VALVE (ID 823045) WAS RETURNED FOR EVALUATION. FAILURE ANALYSIS - THE CATHETER WAS VISUALLY INSPECTED; NO DEFECTS WERE NOTED. THE CATHETER WAS IRRIGATED, NO OCCLUSIONS WERE NOTED. THE VALVE WAS LEAK TESTED; LEAKS WERE NOTED. ROOT CAUSE ANALYSIS ¿ NO ROOT CAUSE COULD BE DETERMINED AS THE TECHNICIAN COULD NOT CONFIRM ANY PROBLEM WITH THE CATHETER AT THE TIME OF INVESTIGATION. A POSSIBLE ROOT CAUSE FOR THE ISSUE REPORTED BY THE CUSTOMER COULD BE DUE TO BIOLOGICAL DEBRIS AND PROTEIN BUILD UP INTERFERING WITH THE CATHETER. AT THE TIME OF INVESTIGATION, NO FUNCTION ISSUES WERE NOTED.

Additional Manufacturer Narrative · 0

AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

N/A.

Description of Event or Problem · 0

A PHYSICIAN REPORTED A HAKIM PERITONEAL CATHETER (ID 823045) AND A CONNECTOR (ID 823053) WERE IMPLANTED IN A 42 YEAR-OLD MALE PATIENT VIA VENTRICULOPERITONEAL SHUNT ON (B)(6) 2023. THE CATHETER WAS CONNECTED TO A COMPETITOR¿S VALVE WHICH WAS ALREADY IMPLANTED. ON (B)(6), 2023, OVERDRAINAGE WAS OBSERVED. A SHUNT CONTRAST WAS PERFORMED AND CEREBROSPINAL FLUID (CSF) LEAKAGE FROM THE CATHETER WAS FOUND. ON (B)(6) 2023, THE CATHETER WAS EXPLANTED AND REPLACED, CONNECTOR REMAINED IMPLANTED. ACCORDING TO INFORMATION PROVIDED, IT IS UNKNOWN IF THE PATIENT EXPERIENCED ANY SIGNS AND SYMPTOMS DUE TO OVERDRAINAGE AND CSF LEAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1744892 HAKIM PERITON CATH,120CM STANDARD CATHETERS JXG INTEGRA LIFESCIENCES MANSFIELD 10381780518839

Patients

Seq Age Sex Outcome Treatment
1 42 YR Male SHUNT CONNECTOR (ID 823053)