FDA Adverse Event Injury Summary report: N

NOA MEDICAL SOFT TOUCH SIDE RAIL

MDR report key: 1818793 · Received August 24, 2010

Report

Report Number
1931947-2010-00001
Event Type
Injury
Date Received
August 24, 2010
Date of Event
August 2, 2010
Report Date
August 3, 2010
Manufacturer
NOA MEDICAL INDUSTRIES, INC.
Product Code
FNL
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS INSPECTED BY A TECHNICIAN FROM THE DISTRIBUTOR/RENTAL AGENCY, (B)(4). (B)(4) SENT A COPY OF THE SIDE RAIL SERVICE MANUAL TO ALL FACILITIES HIGHLIGHTING THE INSTRUCTIONS TO TIGHTEN THE ATTACHMENT PINS WITH A 1/2 OR ADJUSTABLE WRENCH. (B)(4) SENT TECHNICIANS TO ALL FACILITIES TO INSPECT ATTACHMENT PINS FOR PROPER INSTALLATION AND TO TIGHTEN ANY LOOSE PINS.

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT ALLEGEDLY A SIDE RAIL ATTACHMENT PIN HAD BROKEN RESULTING IN THE RESIDENT FALLING AND FRACTURING HIS SHOULDER. A TECHNICIAN WAS SENT TO THE FACILITY AND FOUND THAT THE PIN HAD NOT BROKEN BUT IT WAS LOOSE. THE TECHNICIAN TIGHTENED THE PIN AND PUT THE RAIL BACK INTO SERVICE. THE ATTACHMENT PINS ARE ASSEMBLED ONTO THE SIDE RAILS BY THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NOA MEDICAL SOFT TOUCH SIDE RAIL HOSPITAL BED SIDE RAIL FNL NOA MEDICAL INDUSTRIES, INC. 7540010BEI

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization