FDA Adverse Event Injury Summary report: N

VIPER TI SAI POLY 8X90MM

MDR report key: 18184116 · Received November 21, 2023

Report

Report Number
1526439-2023-02408
Event Type
Injury
Date Received
November 21, 2023
Manufacturer
MEDOS INTERNATIONAL SàRL CH
Product Code
NKB
UDI-DI
10705034265742
PMA / PMN Number
K111571
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. A REVIEW OF THE RECEIVING INSPECTION (RI) FOR VIPER TI SAI POLY 8X90MM WAS CONDUCTED IDENTIFYING THAT LOT NUMBER TBAIIM WAS RELEASED IN ONE BATCH. BATCH1: LOT QTY OF (B)(4) UNITS WERE RELEASED ON 26 JAN 2022 WITH NO DISCREPANCIES. SUPPLIER: DEPUY: (B)(4). AS A RESULT, THE RI IDENTIFIED NO ISSUES DURING THE MANUFACTURING AND RELEASE OF THIS DEVICE THAT COULD HAVE CONTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER. THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL INSPECTION OF THE RETURNED SAMPLE REVEALED THAT THE VIPER TI SAI POLY 8X90MM EXHIBITS SIGNS OF STRIPPED FROM THE INNER THREADED HEAD, THE CONDITION FOUND CANNOT BE DETERMINED WHETHER IT OCCURRED IN THE FIRST SURGERY OR DURING THE REVISION. THERE ARE NO X-RAYS TO CONFIRM THAT THE LOOSE CONDITION HAD PRESENT BETWEEN THE SINGLE-INNER SETSCREW AND THE VIPER TI SAI POLY 8X90MM. A FUNCTIONAL TEST WAS PERFORMED AND DUE TO THE DAMAGE FOUND IN THE INNER THREADED VIPER SCREW HEAD, IT COULD NOT BE THREADED. HOWEVER, THE REPORTED CONDITION CANNOT BE CONFIRMED. A DIMENSIONAL WAS UNABLE TO BE PERFORMED DUE TO POST MANUFACTURING DAMAGE. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE VIPER TI SAI POLY 8X90MM WOULD NOT CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THIS MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H3, H6: THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H10 ADDITIONAL NARRATIVE: D2B: ADDITIONAL PRODUCT CODES: MNI, KWP AND MNH. D9: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. E3: REPORTER IS A J&J EMPLOYEE. H3, H6: THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

PATIENT HAS CLAIMED SYMPTOM, AND THE SURGEON DISCOVERED THE SET SCREW HAS LOOSENED.

Description of Event or Problem · 0

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A SPINAL FUSION ON AN UNKNOWN DATE ON (B)(6) 2023. AFTER THE SURGERY, THE PATIENT HAD SYMPTOMS. THE SECOND SURGERY WAS PERFORMED ON (B)(6) 2023, AND THE SET SCREW IN QUESTION WAS FOUND TO BE DETACHED FROM THE SCREW IN QUESTION. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THE CAUSE IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) VIPER TI SAI POLY 8X90MM. THIS IS REPORT 1 OF 2 FOR COMPLAINT: (B)(6).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2068049 VIPER TI SAI POLY 8X90MM THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM NKB MEDOS INTERNATIONAL SàRL CH TBAIIM 10705034265742

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention SINGLE-INNER SETSCREW.| SINGLE-INNER SETSCREW.| SINGLE-INNER SETSCREW.| SINGLE-INNER SETSCREW.