HF SENSOR DELIVERY SYSTEM
Report
- Report Number
- 3004936110-2023-01440
- Event Type
- Injury
- Date Received
- November 21, 2023
- Date of Event
- October 25, 2023
- Report Date
- January 30, 2024
- Manufacturer
- ST. JUDE MEDICAL, INC.
- Product Code
- MOM
- UDI-DI
- 05414734509237
- PMA / PMN Number
- P100045
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE INFORMATION PROVIDED WAS PERFORMED, AND THE SENSOR WAS WITHIN THE TOLERANCE FOR CM MEAN; THEREFORE, THE REPORTED MALFUNCTION IS UNCONFIRMED. THIS REPORTED EVENT WAS INVESTIGATED FOR POSSIBLE INACCURATE READING. BASED ON THE INFORMATION GIVEN AND BACKEND LOG ANALYSIS, IT WAS CONFIRMED THAT THE DEVICE WAS OPERATING WITHIN THE EXPECTED FREQUENCY RANGE OF 30-37.5 MHZ. THE SENSOR WAS OPERATING AT 34.7, 34.4, AND 34.7 MHZ DURING THE REVIEW OF THE APPLICABLE READING(S). A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.
A RIGHT HEART CATHETERIZATION WAS PERFORMED TO CONFIRM THE VALIDITY OF THE PRESSURE SENSOR READINGS. THE SENSOR WAS RECALIBRATED AND THE MEAN INCREASED WAS BY 3.5 MMHG. READINGS WERE OBSERVED UNDER THE MANUFACTURER'S PATIENT CARE NETWORK DATABASE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1855733 | HF SENSOR DELIVERY SYSTEM | SYSTEM, HEMODYNAMIC, IMPLANTABLE | MOM | ST. JUDE MEDICAL, INC. | CM2000 | 8692554 | 05414734509237 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Male | Hospitalization| R |