BELLAVISTA
Report
- Report Number
- 3004553423-2023-02050
- Event Type
- Malfunction
- Date Received
- November 20, 2023
- Date of Event
- October 27, 2023
- Report Date
- November 20, 2023
- Manufacturer
- VYAIRE MEDICAL
- Product Code
- CBK
- UDI-DI
- 07640149388879
- PMA / PMN Number
- K163127
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DEVICE EVALUATION UPDATE: G3, G6, H2, H6, AND H11. ADDITIONAL INFORMATION: D3 RESULTS OF INVESTIGATION: THE SUSPECT DEVICE WAS NOT RETURNED TO VYAIRE MEDICAL FOR EVALUATION, THEREFORE, A DEFINITIVE ROOT CAUSE COULDN'T BE DETERMINED. PLV ALARMS ARE VISIBLE ON THE LOGS TOGETHER WITH OCCLUSION ALARMS. WITH THE AVAILABLE DATA'S FROM THE LOGS, THE ISSUE APPEARED WITH ADULT PATIENTS. THE SET TIDAL VOLUME WAS NOT ACHIEVABLE WITH THE MAXIMUM ALLOWED AIRWAY PRESSURE (PPEAK UPPER ALARM LIMIT -5MBAR), THEREFORE THE PLV (PRESSURE LIMITED VENTILATION) LUNG PROTECTION SAFETY FEATURE HAS KICKED-IN AS DESIGNED AND THE MACHINE HAS ALARMED ACCORDINGLY ALARM 105 - "VOLUME LOW (PLV PRESSURE-LIMITED)".
AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. HOWEVER, VYAIRE TS RECEIVED LOGS DATED FROM AUG AND FOUND THAT ALL 260/262 ALARMS WERE IN PEDS OR ADULT MODE. THEREFORE, ROOT CAUSE HAS NOT BEEN DETERMINED YET. VYAIRE MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE CUSTOMER REPORTED BELLAVISTA1000E US GIVING A REPEATING AND NUISANCE 260/262 CIRCUIT OCCLUSION ALARMS WHILE ON VERY LOW BIRTHWEIGHT BABIES IN NICU (NEONATAL INTENSIVE CARE UNIT). FURTHERMORE, THERE WAS NO INFORMATION REGARDING PATIENT HARM ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1577308 | BELLAVISTA | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | VYAIRE MEDICAL | BELLAVISTA 1000E US 17,3" VENTILATOR | 07640149388879 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |