FDA Adverse Event Malfunction Summary report: N

BECKMAN GS-6R

MDR report key: 181726 · Received August 3, 1998

Report

Report Number
2912312-1998-00001
Event Type
Malfunction
Date Received
August 3, 1998
Date of Event
July 1, 1998
Report Date
July 27, 1998
Manufacturer
BECKMAN INSTRUMENTS, INC.
Product Code
JQC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

CUSTOMER STATES THAT THEY HAD MODIFIED A PLATE CARRIER FROM A TH-4 ROTOR SO THEY COULD SPIN THE PLATES IN THE GS-6R CENTRIFUGE USING A 3.7 YOKE. THE INCORRECT AND MODIFIED CARRIER WAS THEN ACCIDENTALLY SPUN AT A SPEED OF 5500 RPM, WHEN THEIR INTERNAL PROCEDURE CALLED FOR A SPEED OF 1000 RPM. THE CARRIER FAILED UNDER THIS STRESS AND INTERIOR ROTOR COMPONENTS IMPACTED THE FRONT OF THE CHAMBER CAUSING THE PRIMARY AND SECONDARY LID LATCHES TO FAIL. ROTOR PARTS WERE THEN THROWN OUT OF THE CENTRIFUGE INTO THE WORK AREA. CUSTOMER DID NOT NOTIFY MFR OF THIS ISSUE. THIS EVENT WAS IDENTIFIED BY FIELD SERVICE PERSONNEL VISITING THE USER FACILITY ON A ROUTINE SERVICE CALL ON A DIFFERENT INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN GS-6R CENTRIFUGE JQC BECKMAN INSTRUMENTS, INC. GS-6R NA

Patients

Seq Age Sex Outcome Treatment
1 * Other