ALLIED HEALTHCARE PRODUCTS INC.
Report
- Report Number
- 1924066-2023-00003
- Event Type
- Malfunction
- Date Received
- November 19, 2023
- Date of Event
- January 10, 2023
- Report Date
- July 23, 2024
- Manufacturer
- ALLIED HEALTHCARE PRODUCTS INC.
- Product Code
- BTL
- UDI-DI
- 000267200014
- PMA / PMN Number
- K132021
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- 501
Narratives
**UDI RELATED DATA QAULITY UPDATES ONLY** SEVERAL UPDATES HAVE BEEN MADE TO THIS REPORT. THE FULL UDI HAS BEEN ENTERED, ALONG WITH THE 5 10(K). THE INFORMATION ENTERED IN THIS REPORT AND THAT OF THE GUDID THAT DID NOT MATCH ARE THE BRAND NAME AND THE MANUFACTURER NAME. THE REASONING BEHIND THIS DIFFERENCE IS THAT WE WERE CONSIDERED ALLIED HEALTHCARE PRODUCTS INC. WHEN THE SUSPECT MEDICAL DEVICES WERE MANUFACTURED AS WELL AS WHEN THE MALFUNCTION OCCURRED. THE MEDWATCH WAS SUBMITTED AFTER WE BECAME ALLIED MEDICAL LLC. THE REASON BEHIND THIS IS ALLIED HEALTHCARE PRODUCTS INC. DECLARED BANKRUPTCY AND ITS ASSETS WERE SOLD ON JULY 13, 2023, WHICH WAS WHEN ALLIED MEDICAL LLC WAS CREATED. PLEASE LET ME KNOW IF YOU HAVE ANY QUESTIONS.
THE UNIT WAS TESTED AS RECEIVED AND WAS OUT OF CALIBRATION. THERE WERE NO LEAKS DETECTED AND THE VENT WAS VERIFYED TO BE CHARGING AND THE FAN RUNNING.
PER THE CREW, THIS VENTILATOR SUFFERED A FAILURE DURING TRANSPORT. THEY WERE USING CPAP MODE AND THE UNITL ALARMED AND STOPPED DELIVERING OXYGEN. IT WENT FROM 3/4 BATTERY TO DEAD.
PER THE CREW, THIS VENTILATOR SUFFERED A FAILURE DURING TRANSPORT. THEY WERE USING CPAP MODE AND THE UNIT ALARMED AND STOPPED DELIVERING OXYGEN. IT WENT FROM 3/4 BATTERY TO DEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 916743 | ALLIED HEALTHCARE PRODUCTS INC. | AHP300 | BTL | ALLIED HEALTHCARE PRODUCTS INC. | AHP300 | NO | 000267200014 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |