FDA Adverse Event Malfunction Summary report: N

ALLIED HEALTHCARE PRODUCTS INC.

MDR report key: 18167167 · Received November 19, 2023

Report

Report Number
1924066-2023-00003
Event Type
Malfunction
Date Received
November 19, 2023
Date of Event
January 10, 2023
Report Date
July 23, 2024
Manufacturer
ALLIED HEALTHCARE PRODUCTS INC.
Product Code
BTL
UDI-DI
000267200014
PMA / PMN Number
K132021
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

**UDI RELATED DATA QAULITY UPDATES ONLY** SEVERAL UPDATES HAVE BEEN MADE TO THIS REPORT. THE FULL UDI HAS BEEN ENTERED, ALONG WITH THE 5 10(K). THE INFORMATION ENTERED IN THIS REPORT AND THAT OF THE GUDID THAT DID NOT MATCH ARE THE BRAND NAME AND THE MANUFACTURER NAME. THE REASONING BEHIND THIS DIFFERENCE IS THAT WE WERE CONSIDERED ALLIED HEALTHCARE PRODUCTS INC. WHEN THE SUSPECT MEDICAL DEVICES WERE MANUFACTURED AS WELL AS WHEN THE MALFUNCTION OCCURRED. THE MEDWATCH WAS SUBMITTED AFTER WE BECAME ALLIED MEDICAL LLC. THE REASON BEHIND THIS IS ALLIED HEALTHCARE PRODUCTS INC. DECLARED BANKRUPTCY AND ITS ASSETS WERE SOLD ON JULY 13, 2023, WHICH WAS WHEN ALLIED MEDICAL LLC WAS CREATED. PLEASE LET ME KNOW IF YOU HAVE ANY QUESTIONS.

Additional Manufacturer Narrative · 0

THE UNIT WAS TESTED AS RECEIVED AND WAS OUT OF CALIBRATION. THERE WERE NO LEAKS DETECTED AND THE VENT WAS VERIFYED TO BE CHARGING AND THE FAN RUNNING.

Description of Event or Problem · 0

PER THE CREW, THIS VENTILATOR SUFFERED A FAILURE DURING TRANSPORT. THEY WERE USING CPAP MODE AND THE UNITL ALARMED AND STOPPED DELIVERING OXYGEN. IT WENT FROM 3/4 BATTERY TO DEAD.

Description of Event or Problem · 0

PER THE CREW, THIS VENTILATOR SUFFERED A FAILURE DURING TRANSPORT. THEY WERE USING CPAP MODE AND THE UNIT ALARMED AND STOPPED DELIVERING OXYGEN. IT WENT FROM 3/4 BATTERY TO DEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
916743 ALLIED HEALTHCARE PRODUCTS INC. AHP300 BTL ALLIED HEALTHCARE PRODUCTS INC. AHP300 NO 000267200014

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown