FDA Adverse Event Malfunction Summary report: N

BD NEXIVA

MDR report key: 18166778 · Received November 18, 2023

Report

Report Number
1710034-2023-01319
Event Type
Malfunction
Date Received
November 18, 2023
Date of Event
October 21, 2023
Report Date
January 10, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
30382903835363
PMA / PMN Number
K102520
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4) FOLLOW UP MDR FOR DEVICE EVALUATION: NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 383536 AND LOT NUMBER 3150143. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. H3 OTHER TEXT : SEE MANUFACTURER NARRATIVE.

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NEXIVA LEAKED DURING CONTRAST INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING HIGH PRESSURE CONTRAST, A SMALL HOLE DEVELOPED AND SQUIRTED CONTRAST ON PATIENT AND GIVER. HOSPITAL IS FILING A MEDSUN REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NEXIVA LEAKED DURING CONTRAST INFUSION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING HIGH PRESSURE CONTRAST, A SMALL HOLE DEVELOPED AND SQUIRTED CONTRAST ON PATIENT AND GIVER. HOSPITAL IS FILING A MEDSUN REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673433 BD NEXIVA PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 3150143 30382903835363

Patients

Seq Age Sex Outcome Treatment
1 Unknown