FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 18165494 · Received November 17, 2023

Report

Report Number
2955842-2023-20422
Event Type
Malfunction
Date Received
November 17, 2023
Date of Event
October 31, 2023
Report Date
October 31, 2023
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110898
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED EVENT. THE ISI FSE WAS ABLE TO REPRODUCE THE REPORTED ISSUE AND REPLACED THE INTEGRATED ELECTROSURGICAL UNIT (IESU) TO CORRECT THE REPORTED PROBLEM. THE SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE. THE COMPLAINT WAS CONFIRMED BASED ON THE FIELD EVALUATION. ISI HAS RECEIVED THE PART; HOWEVER, FAILURE ANALYSIS HAS NOT COMPLETED THEIR INVESTIGATION. A FOLLOW-UP MDR WILL BE SUBMITTED WHEN FAILURE ANALYSIS HAS COMPLETED THEIR INVESTIGATION.

Additional Manufacturer Narrative · 0

FAILURE ANALYSIS (FA) HAS COMPLETED THEIR INVESTIGATION ON THE VIO INTEGRATED ELECTRO SURGICAL GENERATOR UNIT (IESU). FA WAS ABLE TO REPRODUCE THE ISSUE BY PLACING THE UNIT ON AN IN-HOUSE SYSTEM AND WAS RUN IN NORMAL MODE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED CHOLECYSTECTOMY SURGICAL PROCEDURE, THE INTUITIVE SURGICAL, INC. (ISI) CLINICAL SALES REPRESENTATIVE (CSR) CONTACTED THE ISI TECHNICAL SUPPORT ENGINEER (TSE) AND REPORTED THAT THE SITE WAS EXPERIENCING C-34 ERROR ON THE INTEGRATED ELECTROSURGICAL UNIT (IESU). THE ISI TSE VIEWED LOGS AND CONFIRMED THE ERROR. THE ISI CSR STATED THAT THE SITE TRIED NEW ENERGY CORDS AND INSTRUMENTS WITH NO CHANGE. THE ISI CSR STATED THAT THE SITE ALSO TRIED TO RESTART THE IESU WITH NO CHANGE. THE PROCEDURE WAS CONVERTED TO LAPAROSCOPIC SURGERY WITH NO REPORTED INJURY.

Description of Event or Problem · 0

REFER TO H10/H11 FOR FOLLOW-UP INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1744569 DA VINCI XI VISION SIDE SYSTEM NAY INTUITIVE SURGICAL, INC 381121-10 N/A 00886874110898

Patients

Seq Age Sex Outcome Treatment
1 Unknown DA VINCI INSTRUMENTS AND ACCESSORIES