CAPIOX FX25 OXYGENATOR
Report
- Report Number
- 9681834-2023-00220
- Event Type
- Malfunction
- Date Received
- November 17, 2023
- Date of Event
- October 23, 2023
- Report Date
- November 17, 2023
- Manufacturer
- TERUMO MEDICAL CORPORATION
- Product Code
- DTZ
- PMA / PMN Number
- K071572
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
G4: PMA/510(K): K130280; E3: OCCUPATION: CLINICAL ENGINEER. VISUAL INSPECTION OF THE THERMISTOR PROBES OF OXYGENATOR AND OF THE RESERVOIR OBTAINED THE FOLLOWING RESULT. THERE WAS NO ANOMALY SUCH AS A BREAKAGE. NO ANOMALY IN THE CONDITION OF ADHESION BETWEEN THE THERMISTOR PROBES AND THE PORTS. THE ACTUAL SAMPLE (AFTER CLEANING) WAS INCORPORATED INTO A CIRCUIT CONSISTING OF TUBING. WHILE SALINE SOLUTION WAS CIRCULATED THROUGH THE BLOOD CHANNEL, 37°C WARM WATER WAS FLOWED INTO THE HEAT EXCHANGER. THE DISPLAYED TEMPERATURE WAS COMPARABLE TO THAT READ WITH A CURRENT PRODUCT AND NO ANOMALY WAS OBSERVED. THE MANUFACTURING RECORD AND THE SHIPPING INSPECTION RECORD OF THE ACTUAL SAMPLE FOUND NO ANOMALY. NO OTHER SIMILAR REPORT OF THE PRODUCT WITH THE INVOLVED PRODUCT CODE/LOT NUMBER WAS FOUND. BASED ON THE INVESTIGATION RESULTS, NO ANOMALY WAS FOUND IN EACH THERMISTOR PROBE OF THE ACTUAL SAMPLE, AND THE CAUSE OF THE OCCURRENCE OF THIS CASE COULD NOT BE CLARIFIED. RELEVANT INSTRUCTIONS FOR USE (IFU) REFERENCE: "DO NOT USE IF THE PACKAGE OR DEVICE IS DAMAGED (E.G. CRACKED) OR ANY OF THE PORT CAPS ARE OFF." TERUMO MEDICAL PRODUCTS (TMP) (IMPORTER) REGISTRATION NO. 2243441 IS SUBMITTING THIS REPORT ON BEHALF OF ASHITAKA FACTORY OF TERUMO CORPORATION (MANUFACTURER) REGISTRATION NO. 9681834.
THE USER FACILITY REPORTED THAT THE ARTERIAL BLOOD TEMPERATURE AND VENOUS BLOOD TEMPERATURE ON THE DISPLAY WERE UNSTABLE AND DISORDERED. THE PROBLEM PERSISTED EVEN AFTER THE THERMISTOR CABLES WERE REPLACED WITH NEW ONES. THE PUMP CHART SHOWED THAT VENOUS BLOOD TEMPERATURE WAS UNSTABLE, INCLUDING FROM 11:16 TO 11:19, AND THE ARTERIAL BLOOD TEMPERATURE WAS OBSERVED UNSTABLE AROUND 14:36. THE THERMISTOR CABLES WERE REPLACED WITH NEW ONES. THE SURGERY CONTINUED AND WAS COMPLETED SUCCESSFULLY. THE FINAL PATIENT IMPACT WAS NOT HARMED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1745286 | CAPIOX FX25 OXYGENATOR | OXYGENATOR, CARDIOPULMONARY BYPASS | DTZ | TERUMO MEDICAL CORPORATION | N/A | 230525 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |