FDA Adverse Event Malfunction Summary report: N

HAMILTON-C6

MDR report key: 18146511 · Received November 16, 2023

Report

Report Number
3001421318-2023-03875
Event Type
Malfunction
Date Received
November 16, 2023
Date of Event
September 4, 2023
Report Date
August 16, 2024
Manufacturer
HAMILTON MEDICAL AG
Product Code
CBK
UDI-DI
07630002808590
PMA / PMN Number
K201658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: FSN WAS INITIATED: Z-0267-2023 (SEPTEMBER 22, 2022): IN VERY RARE CASES, THE BACKLIGHT OF THE HAMILTON-C6 DISPLAY MAY FAIL. THIS IS ATTRIBUTABLE TO A RESET OF THE HMI CONTROLLER (HMI RESET) WHICH CAUSES THE DISPLAY TO FAIL FOR 2-3 SECONDS. THE DEVICE'S VENTILATION FUNCTION IS NOT AFFECTED BY THIS.

Additional Manufacturer Narrative · 0

HAMILTON MEDICAL AG COMES TO THE CONCLUSION: FSN WAS INITIATED: Z-0267-2023 (B)(6) 2022: IN VERY RARE CASES, THE BACKLIGHT OF THE HAMILTON-C6 DISPLAY MAY FAIL. THIS IS ATTRIBUTABLE TO A RESET OF THE HMI CONTROLLER (HMI RESET) WHICH CAUSES THE DISPLAY TO FAIL FOR 2-3 SECONDS. THE DEVICE'S VENTILATION FUNCTION IS NOT AFFECTED BY THIS. HAMILTON MEDICAL AG COMPLAINT NUMBER: CER (B)(4). FOLLOW-UP 1 - CORRECTED INFORMATION: **UDI RELATED DATA QUALITY UPDATES ONLY** FIELD D4 WAS UPDATED WITH FULL UDI INFORMATION AS REQUESTED.

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: INTERACTION PANEL BLACKOUT FOR 2-3 SEC (HMI RESET).

Description of Event or Problem · 0

THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: INTERACTION PANEL BLACKOUT FOR 2-3 SEC (HMI RESET).