FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,SBM,10

MDR report key: 18144367 · Received November 15, 2023

Report

Report Number
0002023141-2023-03256
Event Type
Injury
Date Received
November 15, 2023
Date of Event
June 17, 2023
Report Date
April 3, 2024
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019225
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). D10: CONCOMITANT MEDICAL PRODUCT AND THERAPY DATES TSVB10 X 3 IMPL TAPERED SCR-V SBM 3.7MM 3.5MM 10MM LOT NUMBER 1245286 X 2/1253190, TSV4B10 IMP,TSV,4.1MM,SBM,10 LOT NUMBER 1243780 AND TSVB8 IMPL TAPERED SCR-V SBM 3. 7MM 3.5MM 8MM LOT NUMBER 1253135 . A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). THE DENTAL CENTER HAS CONFIRMED THAT THEY ARE UNABLE TO PROVIDE THE EXACT PLACEMENT DATE. THEREFORE THE PLACEMENT DATE FIELD HAS CHANGED TO UNKNOWN. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1254556. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. STERILIZATION RECORD WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1254556 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: PERI-IMPLANTITIS. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR PERI-IMPLANTITIS EVENTS RECOGNIZING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO A WIDE RANGE OF EXTERNAL FACTORS (NON-DESIGN OR MANUFACTURING RELATED), INCLUDING MEDICAL CONDITIONS (E.G., DIABETES, POOR BONE QUALITY, ETC.) / PATIENT HABITS (E.G., SMOKING) AND SURGICAL TECHNIQUE. PREVIOUSLY COMPLETED INVESTIGATIONS FOR THESE EVENTS HAVE NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. FURTHERMORE, THE PROBABILITY OF A MANUFACTURING OR DESIGN DEFECT THAT MIGHT LEAD TO PERI-IMPLANTITIS OCCURRING AND ESCAPING THE AVAILABLE DETECTIONS HAS BEEN ASSESSED AND FOUND REMOTE AND ALMOST NONEXISTENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

THE DOCTOR REPORTS THAT THE DENTAL IMPLANTS LOCATED IN UNKNOWN DENTAL POSITIONS, FAILED DUE TO PERI-IMPLANTITIS. THE DOCTOR REPORTS IN THE PER THAT THE PROCEDURE WAS NOT CONCLUDED BY PLACING ANOTHER IMPLANTS. THE DOCTOR REPORTED DIABETES.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2238505 IMP,TSV,4.1MM,SBM,10 DENTAL IMPLANT DZE ZIMMER DENTAL 1254556 00889024019225

Patients

Seq Age Sex Outcome Treatment
1 60 YR Male DENTAL IMPLANT.