RADIAL JAW 3 BIOPSY FORCEPS
Report
- Report Number
- 3005099803-2010-03638
- Event Type
- Malfunction
- Date Received
- August 25, 2010
- Date of Event
- May 12, 2010
- Report Date
- August 2, 2010
- Manufacturer
- BOSTON SCIENTIFIC - COSTA RICA
- Product Code
- FCL
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4):THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.(B)(4)
A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND NO ABNORMALITIES AND THE DEVICE WAS WITHIN SPECIFICATIONS. SIMILARLY, THE RETURNED UNIT WAS FUNCTIONALLY TESTED AND IT PERFORMED ACCORDING TO SPECIFICATIONS. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS NOT CONSISTENT WITH THE COMPLAINT; THAT THE JAWS WON'T CLOSE. THE MOST PROBABLE ROOT CAUSE COULD NOT BE CONFIRMED AS THE RETURNED DEVICE WAS WITHIN MANUFACTURING SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A LABELING REVIEW WAS PERFORMED AND NO ANOMALY WAS FOUND. (B)(4).
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS WAS USED DURING A LARGE INTESTINE BIOPSY PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS UNABLE TO OPEN AND CLOSE CORRECTLY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS WAS USED DURING A LARGE INTESTINE BIOPSY PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS UNABLE TO OPEN AND CLOSE CORRECTLY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIAL JAW 3 BIOPSY FORCEPS | FORCEPS, BIOPSY, NON-ELECTRIC | FCL | BOSTON SCIENTIFIC - COSTA RICA | M00515360 | 13228082 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |