FDA Adverse Event Malfunction Summary report: N

RADIAL JAW 3 BIOPSY FORCEPS

MDR report key: 1813616 · Received August 25, 2010

Report

Report Number
3005099803-2010-03638
Event Type
Malfunction
Date Received
August 25, 2010
Date of Event
May 12, 2010
Report Date
August 2, 2010
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA
Product Code
FCL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4):THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.(B)(4)

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND NO ABNORMALITIES AND THE DEVICE WAS WITHIN SPECIFICATIONS. SIMILARLY, THE RETURNED UNIT WAS FUNCTIONALLY TESTED AND IT PERFORMED ACCORDING TO SPECIFICATIONS. THE CONDITION OF THE RETURNED INCIDENT DEVICE WAS NOT CONSISTENT WITH THE COMPLAINT; THAT THE JAWS WON'T CLOSE. THE MOST PROBABLE ROOT CAUSE COULD NOT BE CONFIRMED AS THE RETURNED DEVICE WAS WITHIN MANUFACTURING SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED; NO ANOMALIES WERE NOTED. A LABELING REVIEW WAS PERFORMED AND NO ANOMALY WAS FOUND. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS WAS USED DURING A LARGE INTESTINE BIOPSY PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS UNABLE TO OPEN AND CLOSE CORRECTLY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RADIAL JAW 3 SINGLE-USE BIOPSY FORCEPS WAS USED DURING A LARGE INTESTINE BIOPSY PROCEDURE PERFORMED ON (B)(6), 2010. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE DEVICE WAS UNABLE TO OPEN AND CLOSE CORRECTLY. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION REGARDING THE CIRCUMSTANCES SURROUNDING THIS EVENT HAVE BEEN UNSUCCESSFUL TO DATE. SHOULD ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RADIAL JAW 3 BIOPSY FORCEPS FORCEPS, BIOPSY, NON-ELECTRIC FCL BOSTON SCIENTIFIC - COSTA RICA M00515360 13228082

Patients

Seq Age Sex Outcome Treatment
1