FDA Adverse Event Injury Summary report: N

PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM

MDR report key: 18136001 · Received November 14, 2023

Report

Report Number
2024168-2023-12641
Event Type
Injury
Date Received
November 14, 2023
Date of Event
October 23, 2023
Report Date
February 22, 2024
Manufacturer
ABBOTT VASCULAR
Product Code
MGB
UDI-DI
28717648235188
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL PERCLOSE PROSTYLE DEVICE REFERENCED IN B5 IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Additional Manufacturer Narrative · 0

ANALYSIS WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. A REVIEW OF THE LOT HISTORY RECORD IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.D4 - LOT NUMBER UPDATED FROM 3072441 TO 3073141.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS A VESSEL CLOSURE OF AN UNKNOWN VESSEL USING THE PRE-CLOSE TECHNIQUE PRIOR TO AN ENDOVASCULAR ANEURYSM REPAIR (EVAR) PROCEDURE. REPORTEDLY, TWO PROSTYLE SUTURES DID NOT COME OUT. THE METHOD OF ACHIEVING HEMOSTASIS IS UNSPECIFIED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
906831 PERCLOSE PROSTYLE SUTURE-MEDIATED CLOSURE AND REPAIR SYSTEM SUTURE MEDIATED CLOSURE MGB ABBOTT VASCULAR 3073141 28717648235188

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention