UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN
Report
- Report Number
- 1423500-2010-02852
- Event Type
- Malfunction
- Date Received
- August 24, 2010
- Date of Event
- July 31, 2010
- Report Date
- August 5, 2010
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- KDJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NURSE
Narratives
(B)(4). DEVICE EVALUATION ANTICIPATED, BUT NOT YET RECEIVED. ANY RESULTS OF EVALUATION WILL BE PROVIDED IN A FOLLOW UP E MDR. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME AS OR SIMILAR TO PRODUCT DISTRIBUTED WITHIN THE U.S.
(B)(4). THIS COMPLAINT WAS CONFIRMED IN THE LAB FOR TUBING LEAKAGE DUE TO A CUT/HOLE IN THE TUBING (1 1/2 INCHES FROM THE BASE OF THE WHITE TWIST SLEEVE WHEN CLOSED). EXACT ROOT CAUSE CANNOT BE CONFIRMED FOR THE CUT/SLICE/HOLE IN THE TUBING, HOWEVER, SOME POTENTIAL ROOT CAUSES ARE THE FOLLOWING: PATIENT FOLDING THE TUBE IN THE SAME LOCATION MULTIPLE TIMES. PATIENT MAY INCIDENTALLY CAUSE TUBING DAMAGE BY CONTACTING SHARP OBJECTS (ZIPPER, BUCKLE, OR A BED FENCE). MECHANICAL ASSIST TOOLS SUCH AS A HEMOSTAT CREATES CUTS IN THE TUBING WHEN CLAMPED. PATIENT'S METHOD OF PLACING THE SETS INTO THE UV GERMICIDAL EXCHANGE DEVICE (UV CLEAN FLASH ) INCORRECTLY. RENAL QUALITY ENGINEERING, ALONG WITH PLANT MANUFACTURING AND QUALITY PERSONNEL, WILL CONTINUE TO MONITOR PRODUCT LINES FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.
A CUSTOMER CONTACTED BAXTER'S GLOBAL TECHNICAL SERVICE CENTER TO REPORT A LEAK IN THE TUBING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THE ACTUAL SAMPLE IS AVAILABLE FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN | SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE | KDJ | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |