FDA Adverse Event Malfunction Summary report: N

UV-FLASH SOLUTION TRANSFER SET(SHORT) - JAPAN

MDR report key: 1813481 · Received August 24, 2010

Report

Report Number
1423500-2010-02852
Event Type
Malfunction
Date Received
August 24, 2010
Date of Event
July 31, 2010
Report Date
August 5, 2010
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION ANTICIPATED, BUT NOT YET RECEIVED. ANY RESULTS OF EVALUATION WILL BE PROVIDED IN A FOLLOW UP E MDR. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR AS THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE U.S. AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, IT IS BEING REPORTED BECAUSE IT IS THE SAME AS OR SIMILAR TO PRODUCT DISTRIBUTED WITHIN THE U.S.

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT WAS CONFIRMED IN THE LAB FOR TUBING LEAKAGE DUE TO A CUT/HOLE IN THE TUBING (1 1/2 INCHES FROM THE BASE OF THE WHITE TWIST SLEEVE WHEN CLOSED). EXACT ROOT CAUSE CANNOT BE CONFIRMED FOR THE CUT/SLICE/HOLE IN THE TUBING, HOWEVER, SOME POTENTIAL ROOT CAUSES ARE THE FOLLOWING: PATIENT FOLDING THE TUBE IN THE SAME LOCATION MULTIPLE TIMES. PATIENT MAY INCIDENTALLY CAUSE TUBING DAMAGE BY CONTACTING SHARP OBJECTS (ZIPPER, BUCKLE, OR A BED FENCE). MECHANICAL ASSIST TOOLS SUCH AS A HEMOSTAT CREATES CUTS IN THE TUBING WHEN CLAMPED. PATIENT'S METHOD OF PLACING THE SETS INTO THE UV GERMICIDAL EXCHANGE DEVICE (UV CLEAN FLASH ) INCORRECTLY. RENAL QUALITY ENGINEERING, ALONG WITH PLANT MANUFACTURING AND QUALITY PERSONNEL, WILL CONTINUE TO MONITOR PRODUCT LINES FOR TRENDS AND WILL TAKE CORRECTIVE/PREVENTIVE ACTION AS APPROPRIATE.

Description of Event or Problem · 1

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S GLOBAL TECHNICAL SERVICE CENTER TO REPORT A LEAK IN THE TUBING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION. THE ACTUAL SAMPLE IS AVAILABLE FOR EVALUATION.