FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-522NAP PRDGM INS V2.2 PL EN
MDR report key: 1813426
·
Received August 20, 2010
Report
- Report Number
- 3004209178-2010-82598
- Event Type
- Malfunction
- Date Received
- August 20, 2010
- Date of Event
- August 9, 2010
- Report Date
- August 9, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED. RAN A DISPLACEMENT TEST AND FAILED. INSTRUCTED CUSTOMER REVERT TO BACK UP OF MANUAL INJECTIONS. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAP PRDGM INS V2.2 PL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |