FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522LNAH PRDGM INS CL EN ML

MDR report key: 1813391 · Received August 20, 2010

Report

Report Number
2032227-2010-82392
Event Type
Malfunction
Date Received
August 20, 2010
Date of Event
July 27, 2010
Report Date
July 27, 2010
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522LNAH PRDGM INS CL EN ML INFUSION PUMP MDS MEDTRONIC MINIMED MMT-522LNAH

Patients

Seq Age Sex Outcome Treatment
1 35 YR