INNER SHEATH, FOR 26 FR. OUTER SHEATH
Report
- Report Number
- 9610773-2023-03289
- Event Type
- Malfunction
- Date Received
- November 14, 2023
- Date of Event
- October 26, 2023
- Report Date
- July 19, 2024
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HIH
- UDI-DI
- 04042761029339
- PMA / PMN Number
- K931994
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
G4:K931994/ K931995. TO DATE, THE DEVICE HAS NOT BEEN RETURNED TO OLYMPUS FOR EVALUATION. THE INVESTIGATION IS ONGOING AND FOLLOW UP WITH THE USER FACILITY IS CURRENTLY BEING PERFORMED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR IF ANY ADDITIONAL INFORMATION IS PROVIDED BY THE USER FACILITY.
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. CORRECTION ON FIELDS: G2 (HEALTH PROFESSIONAL WAS INADVERTENTLY UNSELECTED I N THE PREVIOUS REPORT). A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR INSPECTION. BASED ON THE RESULTS OF THE INVESTIGATION, A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED. HOWEVER, IT IS LIKELY THAT THAT THE DAMAGE OF THE INSULATION MATERIAL OF THE SHEATH WAS CAUSED BY IMPROPER HANDLING, APPLICATION OF EXCESSIVE FORCE, MECHANICAL IMPACT LIKE FALL, SHOCK OR SIMILAR STRESS, THERMAL MECHANICAL OVERLOAD. ADDITIONALLY, IT CAN BE ASSUMED THE INSULATION TIP'S DAMAGE WAS CAUSED BY MECHANICAL THERMAL INFLUENCE. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE (IFU): 4 BEFORE USE WARNING INFECTION CONTROL RISK: PROPERLY REPROCESS THE PRODUCT BEFORE FIRST AND EACH SUBSEQUENT USE FOLLOWING THE INSTRUCTIONS IN THIS MANUAL AND IN THE SYSTEM GUIDE ENDOSCOPY. IMPROPER AND/OR INCOMPLETE REPROCESSING CAN CAUSE INFECTION OF THE PATIENT AND/OR MEDICAL PERSONNEL. 4.1 INSPECTION AND TESTING INSPECTING THE PRODUCT: VISUALLY INSPECT THE PRODUCT. MAKE SURE THAT IT HAS: -- NO CORROSION -- NO DENTS -- NO SCRATCHES CERAMIC INSULATION AT DISTAL END: VISUALLY INSPECT THE CERAMIC INSULATION AT THE SHEATH¿S DISTAL END BEFORE EACH USE. DO NOT USE THE INSTRUMENT IN CASE OF DAMAGE (E.G. CRACKS, FRACTURES). WARNING RISK OF INJURY: IMPACT, FALL, SHOCK OR SIMILAR STRESS CAN DAMAGE THE CERAMIC INSULATION AT THE SHEATH¿S DISTAL END. DAMAGED INSTRUMENTS CAN CAUSE INJURIES TO THE PATIENT AND/OR USER. DO NOT USE THE INSTRUMENT IF DAMAGED. DAMAGED PRODUCT: IF THE PRODUCT IS DAMAGED OR DOES NOT FUNCTION PROPERLY, CONTACT AN OLYMPUS REPRESENTATIVE OR AN AUTHORIZED SERVICE CENTER." OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE CUSTOMER REPORTED TO OLYMPUS, THE INNER SHEATH, FOR 26 FR. OUTER SHEATH CERAMIC TIP WAS BROKEN. THERE WERE NO REPORTS OF PATIENT HARM.
THE ISSUE WAS FOUND BEFORE THE DIAGNOSTIC CYSTOSCOPY PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1977866 | INNER SHEATH, FOR 26 FR. OUTER SHEATH | HYSTEROSCOPE (AND ACCESSORIES) | HIH | OLYMPUS WINTER & IBE GMBH | A22040A | 23734 | 04042761029339 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |