FDA Adverse Event Malfunction Summary report: N

COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP

MDR report key: 1813350 · Received August 24, 2010

Report

Report Number
6000001-2010-02718
Event Type
Malfunction
Date Received
August 24, 2010
Date of Event
June 29, 2010
Report Date
July 28, 2010
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING A REVIEW OF THE EVENT HISTORY, IT WAS DISCOVERED THAT FAILURE CODE 12:306:399:0169 OCCURRED ONCE ON (B)(6) 2010 DURING INFUSION WHICH CAUSED AN INTERRUPTION DURING DELIVERY.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN REQUESTED, BUT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION WAS CONFIRMED BUT NOT DUPLICATED. THE ASSIGNABLE CAUSE WAS A SOFTWARE FAILURE. NO REPAIR IS NEEDED FOR THE REPORTED CONDITION AT THIS TIME. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS NOT PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS 5.04.00, CATEGORIZED AS UNREMEDIATED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP THAT HAD AN "ALARM FAILURE REGISTER AND IS ACTIVATE". THE EVENT OCCURRED DURING PATIENT THERAPY. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY UNKNOWN

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT THE GREY COLORED HYDROCANISTER LID OF THE UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER CRACKED. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE CX TRIPLE CHANNEL VOLUMETRIC INFUSION PUMP PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1