FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FREEDOM LITE
MDR report key: 1813271
·
Received August 24, 2010
Report
- Report Number
- 2954323-2010-01159
- Event Type
- Malfunction
- Date Received
- August 24, 2010
- Date of Event
- August 3, 2010
- Report Date
- August 24, 2010
- Product Code
- NBW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PRODUCT HAS BEEN REQUESTED BACK FOR AN INVESTIGATION. A FOLLOW-UP REPORT WILL BE FILED ONCE INVESTIGATION RESULTS ARE AVAILABLE. NOTE: THE METER IS DESIGNED TO DISPLAY READINGS OF 20 MG/DL TO 500 MG/DL. THIS METER DOES NOT SHOW A NUMERIC VALUE GREATER THAN 500 MG/DL. A BLOOD GLUCOSE READING ABOVE 500 MG/DL WILL READ AS A "HI" IN THE ADC METER.
Description of Event or Problem · 1
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER WITHIN 10 MINUTES. RESULTS OF 501 MG/DL AND 131 MG/DL WERE PLOTTED ON THE PARKES ERROR GRID. THE RESULTS FELL INTO THE "C" ZONES SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FREEDOM LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | 0932827 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |