ENDOPOUCH RETRIEVER SPEC BAG
Report
- Report Number
- 3005075853-2010-04477
- Event Type
- Malfunction
- Date Received
- August 6, 2010
- Date of Event
- July 19, 2010
- Report Date
- July 19, 2010
- Manufacturer
- ETHICON ENDO-SURGERY, LLC
- Product Code
- GCJ
- PMA / PMN Number
- K011501
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVAL SUMMARY: THE ANALYSIS RESULTS OF THE POUCH FOUND THAT IT WAS RECEIVED WITH THE SUTURE CUT. DURING EVAL THE BAG WAS NOTED TORN AT THE BOTTOM END (NOT AT THE SEAMS). THE TORN PORTION WAS NOTED STRETCHED. EACH DEVICE IS 100% VISUALLY INSPECTED AT TWO DIFFERENT ASSEMBLY OPERATIONS PRIOR TO SHIPMENT AND DAMAGE OF THIS MAGNITUDE WOULD HAVE BEEN DETECTED DURING THIS INSPECTION. HOWEVER, A POTENTIAL CAUSE OF THE TORN BAG IS ATTEMPT TO REMOVE THE BAG WITH SPECIMEN THROUGH THE TROCAR OR FORCE THE BAG THROUGH THE ACCESS SITE AS THIS MAY LEAD TO BAG RUPTURE AND SPILLAGE OF CONTENTS. THE FOLLOWING OPTIONS ARE RECOMMENDED IN ORDER TO AVOID THIS KIND OF ISSUE: REMOVE THE INSTRUMENT, SPECIMEN BAG, AND TROCAR FROM THE ACCESS SITE (DO NOT PULL THE SLIP KNOT). REMOVE THE INSTRUMENT FROM THE TROCAR, FOLLOWING BY THE SPECIMEN BAG WITH THE TROCAR. REMOVE THE INSTRUMENT, TROCAR AND SPECIMEN BAG SEPARATELY FROM THE ACCESS SITE. IF THE BAG WITH THE SPECIMEN CANNOT BE REMOVED THROUGH THE ACCESS SITE, CAREFULLY ENLARGE THE ACCESS SITE TO FACILITATE EASY BAG REMOVAL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MFG PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE BAG BROKE WHEN THE SURGEON WAS TAKING THE GALL BLADDER OUT OF THE ABDOMEN. THE CONTENTS FELL INTO THE PT. THE CONTENTS WERE RETRIEVED WITH SEVERAL PASSES USING A LAPAROSCOPIC STONE RETRIEVAL INSTRUMENT. THE PROCEDURE WAS PROLONGED BY 15 MINUTES. THE LAPAROSCOPIC STONE RETRIEVAL DEVICE AND ADD'L IRRIGATION AND SUCTION WERE USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPOUCH RETRIEVER SPEC BAG | GCJ LAPAROSCOPE, GENERAL & PLASTIC SURGERY | GCJ | ETHICON ENDO-SURGERY, LLC | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |