FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 1813165 · Received August 18, 2010

Report

Report Number
1824206-2010-09435
Event Type
Malfunction
Date Received
August 18, 2010
Date of Event
July 26, 2010
Report Date
July 26, 2010
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE AC PLUG HAD A LOOSE GROUND PIN. HE REPLACED THE AC PLUG TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE GROUND RESISTANCE READING WAS HIGH WHILE DOING AN ELECTRICAL SAFETY TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC POWERED HOSPITAL BED FNL HILL-ROM INC. 1604

Patients

Seq Age Sex Outcome Treatment
1