FDA Adverse Event Injury Summary report: N

MICRO JEWEL

MDR report key: 181307 · Received July 10, 1998

Report

Report Number
2647346-1998-00108
Event Type
Injury
Date Received
July 10, 1998
Date of Event
March 18, 1998
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
LWS
Removal / Correction Number
NI
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

ICD DELIVERED 1ST CHARGE (15J) & 2ND (34J) BUT IMPEDANCE WAS INCREASED & FAILED TO DELIVER CHARGE ON NEXT 3 SHOCKS. PT EXTERNALLY RESCUED. TESTED AGAIN-IMPEDANCE OK W/DISD, BUT W/ICD, IMPEDANCE >200 OHMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO JEWEL Implant IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC MED REL, INC. 7221E NA

Patients

Seq Age Sex Outcome Treatment
1 000 Other