FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 1813038
·
Received August 13, 2010
Report
- Report Number
- 1811755-2010-00945
- Event Type
- Malfunction
- Date Received
- August 13, 2010
- Date of Event
- July 2, 2010
- Report Date
- July 22, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVAL, BUT THE REPORTED CONDITION OF THE DEVICE OVERHEATING COULD NOT BE CONFIRMED. DURING TESTING, EXCESSIVE NOISE AND VIBRATION WERE COMING FROM THE MOTOR ASSEMBLY. THE ROTOR WAS BALANCED AND THE BEARINGS WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING AT THE MID-POINT TO END OF THE PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.