CORE IMPACTION DRILL

MDR report key: 1813038 · Received August 13, 2010

Report

Report Number
1811755-2010-00945
Event Type
Malfunction
Date Received
August 13, 2010
Date of Event
July 2, 2010
Report Date
July 22, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HANDPIECE WAS RECEIVED AT THE MANUFACTURER FOR EVAL, BUT THE REPORTED CONDITION OF THE DEVICE OVERHEATING COULD NOT BE CONFIRMED. DURING TESTING, EXCESSIVE NOISE AND VIBRATION WERE COMING FROM THE MOTOR ASSEMBLY. THE ROTOR WAS BALANCED AND THE BEARINGS WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE HANDPIECE BEGAN OVERHEATING AT THE MID-POINT TO END OF THE PROCEDURE. THERE WAS NO REPORTED PT OR USER INJURY, AND THE CASE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL BONE CUTTING INSTRUMENT AND ACCESSORIES DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK