FDA Adverse Event
Malfunction
Summary report: N
REPAIR CORE SAG SAW
MDR report key: 1813030
·
Received August 12, 2010
Report
- Report Number
- 1811755-2010-00936
- Event Type
- Malfunction
- Date Received
- August 12, 2010
- Date of Event
- July 27, 2010
- Report Date
- July 27, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBE
- PMA / PMN Number
- K032303
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HANDPIECE WAS RECEIVED AT THE MANUFACTURE FOR EVALUATION, AND THE REPORTED CONDITION OF THE DEVICE OVERHEATING WAS DUPLICATED. BASED ON THE INVESTIGATION DETAILS, THE LIKELY CAUSE WAS A CORRODED MOTOR AND GRITTY BEARINGS. THOSE PARTS WERE EACH REPLACED ALONG WITH THE PRESS PLUG, SAG HEAD, AND OTHER COMPONENTS. SERVICE DID A CLEAN, LUBE, AND ADJUST TO THE DEVICE, AND IT WAS REPAIRED AND RETURNED TO THE CUSTOMER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE HANDPIECE OVERHEATED DURING TESTING AT THE ACCOUNT. THIS DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | REPAIR CORE SAG SAW | SURGICAL INSTRUMENT MOTORS AND ACCESSORIES/ATTAC | HBE | STRYKER INSTRUMENTS KALAMAZOO |