ADJ PIN COLLET 2.0-3.2MM
MDR report key: 1813012
·
Received August 11, 2010
Report
- Report Number
- 1811755-2010-00910
- Event Type
- Malfunction
- Date Received
- August 11, 2010
- Date of Event
- July 14, 2010
- Report Date
- July 14, 2010
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MANUFACTURER AND IT WAS DETERMINED THAT THE ROOT CAUSE WAS THE WEAR AND FLAKING OF THE IMPREGLON COATING. THIS COATING IS USED INSIDE OF THE COLLET TO PREVENT THE PIN FROM STICKING. THIS REPORT WILL BE UPDATED IF ADDITIONAL INFO FROM THE INVESTIGATION IS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE COLLET WOULD NOT ACCEPT LARGE AND SMALL PINS AND THE PINS WOULD OCCASIONALLY STICK. THERE WAS NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ PIN COLLET 2.0-3.2MM | INSTRUMENT, SURGICAL, ORTHOPEDIC, DC-POWERED MOTOR | KIJ | STRYKER INSTRUMENTS KALAMAZOO |