FDA Adverse Event Malfunction Summary report: N

MAESTRO DRILL

MDR report key: 1813007 · Received August 11, 2010

Report

Report Number
1811755-2010-00930
Event Type
Malfunction
Date Received
August 11, 2010
Date of Event
June 29, 2010
Report Date
July 21, 2010
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED BY THE MANUFACTURER AND THE COMPLAINT WAS CONFIRMED. A VISUAL EXAMINATION OF THE DRILL CONFIRMED THAT THE PNEUMATIC HOSE WAS TORN OR CUT. THE HOSE ASSEMBLY WAS REPLACED, AND ADDITIONAL PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DEVICE WAS RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING THE CLEANING PROCESS, A TEAR WAS DISCOVERED IN THE HOSE PORTION OF THE PNEUMATIC DRILL. IT IS UNK HOW THE DAMAGE OCCURRED. NO USER INJURY WAS REPORTED, AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.