HAMILTON MEDICAL AG
Report
- Report Number
- 3001421318-2023-21620
- Event Type
- Malfunction
- Date Received
- November 14, 2023
- Date of Event
- June 28, 2021
- Report Date
- November 8, 2023
- Manufacturer
- HAMILTON MEDICAL AG
- Product Code
- CBK
- UDI-DI
- 00730002856789
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE COMPLAINT HAS BEEN REOPENED AND REVIEWED ACCORDING TO FDA FORM 483 INSPECTIONAL OBSERVATION EMS #2, EOBS2 FROM THE FDA INSPECTION CONDUCTED BETWEEN JULY 17 TO JULY 21, 2022 AT THE EMS AND BONADUZ SITES. A DETAILED INVESTIGATION WAS PERFORMED BY AN EXPERT FROM THE TECHNICAL SERVICE: SINCE THE COMPLAINT IN QUESTION WAS SUBMITTED TO HAMILTON MEDICAL AG ALMOST 2 YEARS AGO, NO ATTEMPTS WILL BE PERFORMED TO OBTAIN ADDITIONAL INFORMATION. NO FURTHER INVESTIGATION OR CORRECTION WILL BE PERFORMED EXCEPT THOSE MENTIONED ABOVE. IN FUTURE HAMILTON MEDICAL AG WILL REPORT AN EVENT SIMILAR TO THIS ISSUE AS IT WILL BE DEEMED A REPORTABLE EVENT. THE ALLEGATION IN THIS COMPLAINT WAS CONFIRMED TO BE A COMPLAINT. WITH THIS INVESTIGATION IT HAS BEEN CONFIRMED THAT THE PRODUCT (FLOW SENSOR WITH PRE-ASSEMBLED TUBES) FAILED TO MEET ITS SPECIFICATIONS AT THE TIME OF THE EVENT WHILE THE VENTILATOR WAS USED WITH A PATIENT. THE ROOT CAUSE WAS DETERMINED TO BE AN ASSEMBLY ISSUE DURING THE MANUFACTURING. IN CONSEQUENCE PRODUCTS WERE REPLACED AND THERE WAS A CHANGE IN THE QUALITY CONTROL DURING THE MANUFACTURING PROCESS. THERE WAS NO REPORTED PATIENT OR USER HARM.
FLOW SENSOR TUBES ON THE 260167 COAXIAL CIRCUIT SET WERE PLACED ON BACKWARDS WHEN REMOVED FROM THE PACKAGING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2203476 | HAMILTON MEDICAL AG | FLOW SENSOR SINGLE USE | CBK | HAMILTON MEDICAL AG | FLOW SENSOR SINGLE USE | 00730002856789 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |