FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 1813000 · Received August 6, 2010

Report

Report Number
3005075853-2010-04480
Event Type
Malfunction
Date Received
August 6, 2010
Date of Event
July 1, 2010
Report Date
July 15, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WOULD FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS. THE ANALYSIS RESULTS FOR THE EL5ML DEVICE (B) FOUND THAT IT WAS RETURNED WITH THE SHAFT BENT AT THE MIDDLE. NO FUNCTIONAL TESTING COULD BE PERFORMED DUE TO THE RETURNED CONDITION OF THE DEVICE. POSSIBLE CAUSES FOR THE DAMAGE FOUND MAY BE INADVERTENT PRESSURE PLACED ON THE DEVICE SHAFT. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW. (B)(4), EXPIRATION DATE: 01/2015. MANUFACTURING DATE: (B)(4) 2010. DAMAGED SHAFT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, BOTH DEVICES WOULD NOT RELEASE AT FIRST, EVENTUALLY THE DEVICES RELEASE WHEN THE SURGEON PULLED ON THE HANDLE. A THIRD DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIGAMAX-5MM ENDO CLIP APPLIER FZP CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC NA G4RF7R

Patients

Seq Age Sex Outcome Treatment
1