FDA Adverse Event Malfunction Summary report: N

PUMP MMT-522NAB PRDGM INS V2.2 BL EN

MDR report key: 1812982 · Received August 16, 2010

Report

Report Number
3004209178-2010-82486
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
July 22, 2010
Report Date
July 22, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF MOISTURE DAMAGE ON THE ELECTRONIC AND MOTOR ASSEMBLIES. FURTHER TESTING WAS NOT PERFORMED DUE TO THE FROZEN SCREEN.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP FELL INTO THE SEA BECAUSE THE BELT CLIP BROKE. THE CUSTOMER STATED THAT A LOT OF MOISTURE TRAPPED UNDER THE DISPLAY WAS OBSERVED, AND ALSO THE DEVICE ALARMED. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-522NAB PRDGM INS V2.2 BL EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-522NAB

Patients

Seq Age Sex Outcome Treatment
1