FDA Adverse Event
Malfunction
Summary report: N
PUMP MMT-522NAB PRDGM INS V2.2 BL EN
MDR report key: 1812982
·
Received August 16, 2010
Report
- Report Number
- 3004209178-2010-82486
- Event Type
- Malfunction
- Date Received
- August 16, 2010
- Date of Event
- July 22, 2010
- Report Date
- July 22, 2010
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP WAS RECEIVED WITH A FROZEN DISPLAY AS A RESULT OF MOISTURE DAMAGE ON THE ELECTRONIC AND MOTOR ASSEMBLIES. FURTHER TESTING WAS NOT PERFORMED DUE TO THE FROZEN SCREEN.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP FELL INTO THE SEA BECAUSE THE BELT CLIP BROKE. THE CUSTOMER STATED THAT A LOT OF MOISTURE TRAPPED UNDER THE DISPLAY WAS OBSERVED, AND ALSO THE DEVICE ALARMED. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PUMP MMT-522NAB PRDGM INS V2.2 BL EN | INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-522NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |