PUMP MMT-722NAS PRDGM INS V2.2 SK EN

MDR report key: 1812967 · Received August 16, 2010

Report

Report Number
3004209178-2010-82482
Event Type
Malfunction
Date Received
August 16, 2010
Date of Event
July 12, 2010
Report Date
July 12, 2010
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP FAILED THE DISPLACEMENT TEST AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE. HOWEVER, THE DEVICE PASSED THE PRIME, BASIC OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TESTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HAVING A NO DELIVERY ALARM DURING A MANUAL PRIME. TROUBLESHOOTING WAS DECLINED. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PUMP MMT-722NAS PRDGM INS V2.2 SK EN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAS