FDA Adverse Event
Malfunction
Summary report: N
SOFTLAB
MDR report key: 1812922
·
Received August 16, 2010
Report
- Report Number
- 1058332-2010-00001
- Event Type
- Malfunction
- Date Received
- August 16, 2010
- Report Date
- April 6, 2010
- Manufacturer
- SCC SOFT COMPUTER
- Product Code
- JQP
- Removal / Correction Number
- 1058332-08/06/2010-006-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
AFFECTED PRODUCTS INCLUDE SOFTLAB VERSIONS 4.0.1.13.3, 4.0.1.8 - 4.0.1.14, 4.0.2.0 - 4.0.2.10, 4.0.3.0 - 4.0.3.7. THESE VERSIONS OF SOFTLAB ARE AFFECTED WHEN USED WITH THE SOFTBANK MODULE. METHOD: INSPECTED SOURCE CODE OF PROGRAM.
Description of Event or Problem · 1
SOFTLAB ORDER ENTRY DOES NOT USE THE CORRECT TEST DEFINITION SEARCH TO FIND THE BLOOD PRODUCT WHEN MANUALLY ORDERED. THE TEST DEFINITION SEARCH UTILIZED WAS NOT DEFINED ON THE SYSTEM. WHEN A LIKE-PRODUCT IS ORDERED ON AN EXISTING ORDER USING SOFTLAB ORDER ENTRY, THE CORRESPONDING CROSSMATCH IS NOT ORDERED. A CONFIGURATION CHANGE IN SOFTLAB WILL BE IMPLEMENTED TO REMEDY THE PROBLEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SOFTLAB | JQP | SCC SOFT COMPUTER |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |