FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 1812879 · Received August 12, 2010

Report

Report Number
1824206-2010-09290
Event Type
Malfunction
Date Received
August 12, 2010
Date of Event
July 20, 2010
Report Date
July 20, 2010
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE 22-POSITION CONNECTOR TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THERE IS NO SCALE DISPLAY DUE TO CORROSION ON THE 22-POSITION CONNECTOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED AC POWERED HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1