Additional Manufacturer Narrative · 1
THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED TO ANIMAS. EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. REVIEW OF THE PUMP HISTORY REVEALS SEVERAL LOSS OF PRIME ALARMS ASSOCIATED WITH ZERO FORCE. THE PUMP WAS PRIMED SUCCESSFULLY AND EXERCISED WITH NO LOSS OF PRIME ALARMS DUPLICATED.
Description of Event or Problem · 1
EVALUATION REVEALED A MISALIGNED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS.