FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE
MDR report key: 1812834
·
Received August 12, 2010
Report
- Report Number
- 1824206-2010-09295
- Event Type
- Malfunction
- Date Received
- August 12, 2010
- Date of Event
- July 20, 2010
- Report Date
- July 20, 2010
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE HEAD CYLINDER WAS LEAKING. THE TECHNICIAN REPLACED THE CYLINDER AND CALIBRATED THE POSITION SENSORS TO REPAIR THE BED.
Description of Event or Problem · 1
INFORMATION RECEIVED INDICATES THE HEAD WILL NOT RISE NOR WILL THE BED GO INTO THE CHAIR POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE | AC POWERED HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |