BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC)
Report
- Report Number
- 2647876-2023-00469
- Event Type
- Malfunction
- Date Received
- November 13, 2023
- Date of Event
- October 21, 2023
- Report Date
- December 18, 2023
- Manufacturer
- BECTON DICKINSON CARIBE LTD.
- Product Code
- MDB
- UDI-DI
- 00382904420208
- PMA / PMN Number
- K151866
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H3. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H.6. INVESTIGATION SUMMARY: CATALOG 442020. BATCH NO. 3123425. CUSTOMER REPORTED A FALSE NEGATIVE RESULT. NEITHER PHOTOS NOR RETURNED GOOD SAMPLES WERE RECEIVED. BD WAS UNABLE TO REPRODUCE CUSTOMER¿S EXPERIENCE WITH BACTEC PRODUCT. SATISFACTORY RESULTS WERE OBTAINED FROM RETENTION SAMPLES WHEN TESTED FOR MICROBIAL INSTRUMENT DETECTION AS PER QUALITY PROCEDURES. BATCH HISTORY RECORD REVIEW DID NOT IDENTIFY ANY EVIDENCE FOR WHICH THE CUSTOMER SUBMITTED THE COMPLAINT (I.E. FALSE NEGATIVE). MICROBIAL INSTRUMENT DETECTION INDICATED THAT THE PRODUCT IS PERFORMING AS EXPECTED. A COMPLAINT HISTORY REVIEW WAS CONDUCTED AND ONLY THE CURRENT COMPLAINT WAS FOUND RELATING TO THE INCIDENT LOT NUMBER AND THE ¿AS REPORTED¿ DEFECT CODE. COMPLAINT IS UNCONFIRMED BASED ON RETENTION SAMPLES AND BATCH RECORD REVIEW RESULTS. NO CORRECTIVE ACTIONS WERE REQUIRED. A CROSS FUNCTIONAL TEAM CONTINUALLY MONITORS ALL PRODUCT COMPLAINTS FOR TRENDS AND DETERMINES IF ANY ADDITIONAL ACTIONS ARE NECESSARY BEYOND THE CURRENT INVESTIGATIONAL PROCESS. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT DURING USE WITH THE BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC), A FALSE NEGATIVE RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
IT WAS REPORTED THAT DURING USE WITH THE BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC), A FALSE NEGATIVE RESULT WAS OBTAINED. THERE WAS NO REPORT OF PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1988595 | BD BACTEC¿ PEDS PLUS¿/ F CULTURE VIALS (PLASTIC) | SYSTEM, BLOOD CULTURING | MDB | BECTON DICKINSON CARIBE LTD. | 3123425 | 00382904420208 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |