FDA Adverse Event Malfunction Summary report: N

ECHELON 60

MDR report key: 1812768 · Received August 18, 2010

Report

Report Number
3005075853-2010-04695
Event Type
Malfunction
Date Received
August 18, 2010
Date of Event
July 22, 2010
Report Date
August 4, 2010
Manufacturer
ETHICON ENDO-SURGERY, LLC
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE (B)(4) DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH TWO CARTRIDGE RELOADS PRESENT. RELOAD (C) WAS UNFIRED. RELOAD (B) WAS RECEIVED PARTIALLY FIRED 1/10. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED CARTRIDGE RELOADS WERE TESTED FOR FUNCTIONALITY BY RESETTING AND RELOADING RELOAD (B) INTO THE DEVICE. THE FUNCTIONAL TEST DEMONSTRATED THAT THE REMAINING STAPLES IN THE CARTRIDGE RELOAD FORMED PROPERLY AND THE INSTRUMENT ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED. A BATCH RECORD REVIEW WAS PERFORMED AND NO ANOMALIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED TO BAXTER (B)(4) THAT AT LEAST TWENTY INTERMATE DEVICES WERE LEAKING FROM THE WINGED LUER CAP BEFORE USE. THIS IS REPORT NUMBER 20 OF 20. THE DEVICES WERE FILLED WITH PAMIDRONATE WHEN THE LEAKS WERE OBSERVED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LOW ANTERIOR RESECTION PROCEDURE, THE FIRST RELOAD CAN'T FIRE PROPERLY AND GOT STUCK. THE NURSE CHANGED A NEW DEVICE TO FINISH THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON 60 GDW STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC NA C4RU5E

Patients

Seq Age Sex Outcome Treatment
1 ECR60G, LOT# G4RP6J